VIDAS在保健食品沙门氏菌快速检测中的应用 被引量:4

Fast detection of Salmonella in health food by VIDAS method
收藏 分享 导出
摘要 目的将全自动免疫分析仪(VIDAS)应用于保健食品的沙门氏菌快速检验。方法将21批次保健食品分为BPW组和BPW阳性对照组,分别采用VIDAS和国标方法进行沙门氏菌检验,并比较结果。结果 21批保健食品的沙门氏菌检验中,BPW组两种检验方法结果均为阴性;BPW阳性对照组有20批的VIDAS方法与GB方法的检验结果一致,占总批数的95.2%。结论可以用VIDAS方法对保健食品进行沙门氏菌的快速检验。用国标方法检验沙门氏菌需要96 h或者更长,用VIDAS方法则可以把检验时间缩短到48 h。如果VIDAS方法检验结果为阴性则可以作为最终结果,如果结果为阳性,则需要按照国标方法进一步进行确认试验。 ABSTRACT:Objective The automated immunoassay analyzer (VIDAS) was applied to detect the Salmonella in health food.Methods The 21 lots of health food were divided into BPW group and BPW positive control group. TheSalmonella in 21 lots of health foods were detected by both VIDAS method and GB method. The results were compared.Results All the results were negative in BPW group. Results of both VIDAS method and GB method in 20 lots of health food were consistent with each other in BPW positive control group, and the percentage accounted for 95.2% of the total.Conclusion VIDAS method could be used to fast detect theSalmonella in health food. It took 96 h at least to detectSalmonella by GB method. However, it took only 48 h by VIDAS method. It could be the final result if the result was negative, and it should be confirmed according to the GB method if the result was positive.
作者 王似锦 高春 WANG Si-Jin;GAO Chun;Beijing Institute for Drug Control (Beijing Center for Health Food and Cosmetics Control);
出处 《食品安全质量检测学报》 2015年第2期677-681,共5页 Journal of Food Safety & Quality
关键词 保健食品 沙门氏菌 全自动免疫分析仪 快速检测 health food Salmonella automated immunoassay analyzer fast detection
分类号 TS218 O657
  • 相关文献

参考文献14

二级参考文献131

共引文献177

同被引文献40

引证文献4

二级引证文献7

投稿分析

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部 意见反馈